6JQM
Structure of PaaZ with NADPH
ELECTRON MICROSCOPY
Sample |
---|
PaaZ |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | FEI VITROBOT MARK IV |
Cryogen Name | ETHANE |
Sample Vitrification Details | blotting force 10, blotting time 4 sec, waiting time 15 sec, drying time 0, blotting times 1. |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 103298 |
Reported Resolution (Å) | 3.3 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C3 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 | ||||
Refinement Space | REAL | ||||
Refinement Protocol | OTHER | ||||
Refinement Target | |||||
Overall B Value | 68.9 | ||||
Fitting Procedure | |||||
Details | Real space refinement with secondary structure enabled, minimization and adp |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | FEI FALCON III (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 27 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | 2200 |
Maximum Defocus (nm) | 3200 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2.7 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
Nominal Magnification | 75000 |
Calibrated Magnification | 132075 |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details | Data was collected with EPU software |
EM Software | ||
---|---|---|
Task | Software Package | Version |
PARTICLE SELECTION | Gautomatch | 0.56 |
IMAGE ACQUISITION | EPU | 1.7 |
CTF CORRECTION | Gctf | 1.06 |
MODEL FITTING | Coot | 0.8 |
MODEL REFINEMENT | PHENIX | 1.13 |
INITIAL EULER ASSIGNMENT | RELION | 2.0 |
FINAL EULER ASSIGNMENT | RELION | 2.0 |
CLASSIFICATION | RELION | 2.0 |
RECONSTRUCTION | RELION | 2.0 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
NONE | Particles were automatically corrected for CFT by Relion during data processing. | 175482 |